Alcohol and Alcoholism Advance Access originally published online on October 17, 2007
Alcohol and Alcoholism 2008 43(1):46-48; doi:10.1093/alcalc/agm153
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Evaluation of a new immunoassay for urinary ethyl glucuronide testing
1 Arztpraxis für Medizinische Mikrobiologie und Labordiagnostik, Dessau, Germany
2 Department of Medicine, Karolinska Institute and Karolinska University Hospital, Stockholm, Sweden
3 Department of Clinical Neuroscience, Karolinska Institute and Karolinska University Hospital, Stockholm, Sweden
* Author to whom correspondence should be addressed at: Alcohol Laboratory, L7:03, Karolinska University Hospital Solna, SE-171 76 Stockholm, Sweden. Tel: +46-8-51771530; Fax: +46-8-51771532; Anders.Helander{at}ki.se
Received 29 May 2007; first review notified 17 July 2007; in revised form 3 August 2007; accepted 7 September 2007
| ABSTRACT |
|---|
|
|
|---|
Aims: The minor ethanol metabolite ethyl glucuronide (EtG) is used as a sensitive and specific test for recent alcohol consumption with clinical and forensic applications. This study evaluated a new enzyme immunoassay (DRI-EtG EIA, Microgenics Corp.) for determination of the EtG concentration in urine samples. Methods: Evaluation was done using the kit calibrators (range 0–5.0 mg/L) and controls, an external quality control sample, and 400 consecutive urines from the routine samples pool. The measuring range was extended by dilution of urine samples with saline. Comparison was made with an established liquid chromatographic-mass spectrometric (LC-MS) method. Results: The intra- and inter-assay imprecision of the DRI-EtG EIA in the range 0.4–2.5 mg/L was <2.2% (coefficient of variation, CV), and the limit of quantification was <0.1 mg/L. For the 400 urine samples, the EtG concentrations obtained using the DRI-EtG EIA (mean 24.2 mg/L, range 0–830) and LC-MS method (mean 22.4 mg/L, range 0–959) showed an overall good and statistically significant agreement (r2 = 0.931, P < 0.0001). Conclusions: These results indicated a high level of accuracy and selectivity of the DRI-EtG EIA for quantification of urinary EtG. In the absence of a commonly accepted cut-off limit for urinary EtG, a threshold of 0.5 mg/L (2.2 µmol/L) is proposed, to obtain a high sensitivity but avoid positive results due to unintentional ethanol exposure.
| Introduction |
|---|
|
|
|---|
Ethyl glucuronide (EtG) is a direct phase-II metabolite of ethanol formed by action of UDP-glucuronosyltransferase (UGT) (Foti and Fisher, 2005
1–4 days in a dose-dependent manner (Schmitt et al., 1997
The development of accurate liquid chromatographic-mass spectrometric (LC-MS) methods for qualitative and quantitative determination of EtG (Stephanson et al., 2002
; Weinmann et al., 2004
; Bicker et al., 2006
) allowed for introduction of urinary EtG as a routine test for detection of recent alcohol intake and for monitoring of abstinence. However, LC-MS methods are costly and only available in specialized laboratories, which have hindered a widespread clinical use of this test. A preliminary immunochemical screening method (ELISA) based on polyclonal antibodies was found not to be sensitive and specific enough (Zimmer et al., 2002
), but another more promising prototype enzyme immunoassay (EIA) based on a monoclonal antibody was recently presented (Bih et al., 2006
).
This study evaluated a new and potentially commercially available version of the EIA method, based on a new monoclonal antibody (DRI Ethyl Glucuronide Enzyme Immunoassay [DRI-EtG EIA]; Microgenics Corp.), for analysis of EtG in urine samples. Comparison was made with an established LC-MS method (Stephanson et al., 2002
).
| Materials and Methods |
|---|
|
|
|---|
The urine specimens used for the method comparison were surplus volumes of 400 consecutive de-identified samples from the routine pool at the Alcohol Laboratory, Karolinska University Hospital, Stockholm, for which the EtG concentration had been determined by an electrospray LC-MS method with selected ion monitoring (Stephanson et al., 2002
0.1 mg/L (limit of detection
0.05 mg/L) but for routine clinical purposes a cut-off limit of 0.5 mg/L is applied. The samples were stored frozen at –20°C where the EtG concentration is stable for long time (Stephanson et al., 2002The urine samples were shipped frozen to the Arztpraxis für Medizinische Mikrobiologie und Labordiagnostik in Dessau, Germany for evaluation of the DRI-EtG EIA on an Olympus AU640 instrument. The calibrators (range 0–5.0 mg/L [0–22.5 µmol/L] EtG in urine matrix), controls (0.38–1.25 mg/L in urine matrix), and instrumental application provided by the manufacturer were used together with an external quality control urine sample (EtG target range 1.90–3.50 mg/L) obtained from Medichem (Steinenbronn, Germany). The measuring range of the immunoassay was extended by dilution of urine samples with saline.
|
Owing to a known limitation of the Olympus instrument, urine samples with an optical density >2.5 at the 340 nm measuring wavelength of the DRI-EtG assay (N = 56; 14%) had to be diluted with saline (range 2–10 times, but typically 2 or 4 times) prior to analysis.
| Results and Discussion |
|---|
|
|
|---|
The intra-assay imprecision, expressed as the coefficient of variation (CV; N = 10), for the kit controls was <1.7% and the bias from target values ranged from –8.1 to 8.0% (Table 1). The inter-assay CV (30 measurements on 5 separate days) was <2.2% with a mean bias ranging from –7 to 9.3%. The assay imprecision data demonstrated that the LOQ of the DRI-EtG EIA was <0.1 mg/L, and the test was able to detect 0.1 mg/L with good reproducibility. The intra- and inter-assay CV for the external EtG quality control sample was <1.8% with a bias of <9% from the target (Table 1).
The EtG concentrations obtained using the DRI-EtG EIA (mean 24.2 mg/L, range 0–830) and LC-MS (mean 22.4 mg/L, range 0–959) methods showed an overall good and statistically significant agreement (r2 = 0.931, P < 0.0001, N = 400). For 126 urine samples with EtG values ranging from 0.10 to 50 mg/L by LC-MS method (i.e. using a maximum ten-fold dilution to get into the calibration range of the immunoassay), the r2 was 0.961 (P < 0.0001), the intercept close to the origin and the slope close to 1, and only few outliers were identified (Fig. 1). The good overall quantitative agreement of the DRI-EtG EIA with the LC-MS results indicated a low cross-reactivity of the EtG antibody to other urinary constituents.
|
Two different cut-off limits for urinary EtG (0.5 and 1.0 mg/L, corresponding to 2.2 and 4.5 µmol/L) were evaluated for use with the DRI-EtG EIA. The agreement of qualitative results with the LC-MS method (reference) was 98.2% at both cut-offs (Fig. 2). At the 0.5 mg/L limit, the number of classified positive samples was 8% higher than at the 1.0 mg/L limit. Most of the differing results were due to a combination of assay imprecision and a concentration close to the applied threshold (Fig. 2). However, out of the 400 samples 5 urine samples (1.2%) with clearly deviating EtG results were identified. Two of these contained EtG levels between 0.5 and 1.0 mg/L according to the DRI-EtG EIA, while EtG was not detected by LC-MS method (< 0.05 mg/L). The remaining three samples contained 3.7–9.6 mg/L EtG according to the EIA but concentrations below the clinical cut-off (< 0.5 mg/L) by LC-MS method. All five samples that were positive for EtG by the EIA, but showed negative results or values below the clinical cut-off by LC-MS method, were positive for EtS although showing an unusually low EtG/EtS ratio (Dahl et al., 2002
|
This evaluation demonstrated that the new DRI-EtG EIA is a sensitive and specific method for quantification of EtG in urine samples, and it also offers a low and clinically relevant measuring range (Stephanson et al., 2002
| Conflict of Interest Declaration |
|---|
|
|
|---|
The DRI-EtG EIA kits were provided by Microgenics Corp. but all other costs were covered by the participating laboratories. The authors had complete independence in the interpretation of data and writing of the report.
| References |
|---|
|
|
|---|
Bergström J., Helander A., Jones A. W. Ethyl glucuronide concentrations in two successive urinary voids from drinking drivers: relationship to creatinine content and blood and urine ethanol concentrations. Forensic Science International (2003) 133:86–94.[CrossRef][Web of Science][Medline]
Bicker W., Lammerhofer M., Keller T., et al. Validated method for the determination of the ethanol consumption markers ethyl glucuronide, ethyl phosphate, and ethyl sulfate in human urine by reversed-phase/weak anion exchange liquid chromatography-tandem mass spectrometry. Analytical Chemistry (2006) 78:5884–5892.[Medline]
Bih C., Mitra S., Bodepudi V., et al. Development of a homogenous enzyme immunoassay for the detection of ethyl glucuronide in urine and its evaluation on the MGC 240 analyzer [abstract]. Journal of Analytical Toxicology (2006) 30:146.
Borucki K., Schreiner R., Dierkes J., et al. Detection of recent ethanol intake with new markers: comparison of fatty acid ethyl esters in serum and of ethyl glucuronide and the ratio of 5-hydroxytryptophol to 5-hydroxyindole acetic acid in urine. Alcoholism: Clinical and Experimental Research (2005) 29:781–787.[CrossRef][Web of Science][Medline]
Costantino A., Digregorio E. J., Korn W., et al. The effect of the use of mouthwash on ethylglucuronide concentrations in urine. Journal of Analytical Toxicology (2006) 30:659–662.[Web of Science][Medline]
Dahl H., Stephanson N., Beck O., et al. Comparison of urinary excretion characteristics of ethanol and ethyl glucuronide. Journal of Analytical Toxicology (2002) 26:201–204.[Web of Science][Medline]
Foti R. S., Fisher M. B. Assessment of UDP-glucuronosyltransferase catalyzed formation of ethyl glucuronide in human liver microsomes and recombinant UGTs. Forensic Science International (2005) 153:109–116.[CrossRef][Web of Science][Medline]
Helander A., Beck O. Mass spectrometric identification of ethyl sulfate as an ethanol metabolite in humans. Clinical Chemistry (2004) 50:936–937.
Helander A., Beck O. Ethyl sulfate: a metabolite of ethanol in humans and a potential biomarker of acute alcohol intake. Journal of Analytical Toxicology (2005) 29:270–274.[Web of Science][Medline]
Hoiseth G., Bernard J. P., Karinen R., et al. A pharmacokinetic study of ethyl glucuronide in blood and urine: Applications to forensic toxicology. Forensic Science International (2007) 172:119–124.[CrossRef][Web of Science][Medline]
Jones A. W. Urine as a biological specimen for forensic analysis of alcohol and variability in the urine-to-blood relationship. Toxicological Reviews (2006) 25:15–35.[CrossRef][Medline]
Rohrig T. P., Huber C., Goodson L., et al. Detection of ethylglucuronide in urine following the application of Germ-X. Journal of Analytical Toxicology (2006) 30:703–704.[Web of Science][Medline]
Schmitt G., Aderjan R., Keller T., et al. Ethyl glucuronide: an unusual ethanol metabolite in humans. Synthesis, analytical data, and determination in serum and urine. Journal of Analytical Toxicology (1995) 19:91–94.[Web of Science][Medline]
Schmitt G., Droenner P., Skopp G., et al. Ethyl glucuronide concentration in serum of human volunteers, teetotalers, and suspected drinking drivers. Journal of Forensic Sciences (1997) 42:1099–1102.[Web of Science][Medline]
Stephanson N., Dahl H., Helander A., et al. Direct quantification of ethyl glucuronide in clinical urine samples by liquid chromatography-mass spectrometry. Therapeutic Drug Monitoring (2002) 24:645–651.[CrossRef][Web of Science][Medline]
Weinmann W., Schaefer P., Thierauf A., et al. Confirmatory analysis of ethylglucuronide in urine by liquid-chromatography/electrospray ionization/tandem mass spectrometry according to forensic guidelines. Journal of the American Society of Mass Spectrometry (2004) 15:188–193.[CrossRef]
Wurst F. M., Seidl S., Ladewig D., et al. Ethyl glucuronide: on the time course of excretion in urine during detoxification. Addiction Biology (2002) 7:427–434.[CrossRef][Web of Science][Medline]
Wurst F. M., Skipper G. E., Weinmann W. Ethyl glucuronide–the direct ethanol metabolite on the threshold from science to routine use. Addiction (2003) 98(Suppl 2):51–61.[CrossRef][Web of Science][Medline]
Zimmer H., Schmitt G., Aderjan R. Preliminary immunochemical test for the determination of ethyl glucuronide in serum and urine: comparison of screening method results with gas chromatography-mass spectrometry. Journal of Analytical Toxicology (2002) 26:11–16.[Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Schroder, R. de la Torre, R. Estruch, D. Corella, M. A. Martinez-Gonzalez, J. Salas-Salvado, E. Ros, F. Aros, G. Flores, E. Civit, et al. Alcohol consumption is associated with high concentrations of urinary hydroxytyrosol Am. J. Clinical Nutrition, November 1, 2009; 90(5): 1329 - 1335. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Helander, M. Bottcher, C. Fehr, N. Dahmen, and O. Beck Detection Times for Urinary Ethyl Glucuronide and Ethyl Sulfate in Heavy Drinkers during Alcohol Detoxification Alcohol Alcohol., January 1, 2009; 44(1): 55 - 61. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||